ROD-L M120DC
MODEL M120DC
DC Hipot Test Instrument
Safe, Fast, Efficient
- Digital Display (optional)
- Fully programmable test parameters
- Performs DC dielectric withstand test up to 5.5 KVDC at 10 milliamperes (5mA standard)
- Performs concurrent ground continuity test
- ARC and OVERCURRENT detection circuits
- Fast HV shutdown within 2 milliseconds
- Fast discharge of Device Under Test (DUT) upon shutdown
- Full compliance with UL, VDE, BSI, IEC, CSA, and other test standards
Production/Quality Assurance Testing
- Fully automatic test cycle
- Detects and indicates the following
- Test Ready (security ground chassis sensed)
- Test in Progress
- Test Passed
- Test Failed
- Test parameters easily preset via front panel controls including:
- Voltage ramp
- Test time
- Test voltage
- Overcurrent trip point
- Switch-selectable ramp and rest time combinations
- Arc detection circuit
- Overcurrent detection circuit
- Solid state HV generation and control system for reliability
- Switchable “burn” featured permits destructive testing by maintaining output voltage, following failure, to the DUT throughout duration of preset test time
- Can be interconnected with M25 Ground Continuity Test Instrument or any ROD-L AC Hipot Test Instrument to simplify multiple testing requirements
For specifications, call ROD-L or download the PDF file and refer to page 2.
![]()

Continuos Ground Continuity Monitoring
- Monitors Ground Continuity of DUT before and during each Hipot Test
- Senses for 0.5 ohm (standard), at approximately 1.5 VAC
- Loss of safety ground connection during test causes high voltage to shut down with both audible and visual failure indication
- Inhibits application of high voltage if the security chassis ground is not secure
Superior Safety Features
- Shutdown within 2 milliseconds after dielectric breakdown, overcurrent, or ground continuity fault
- Electronically controlled output voltage ramp rate circuit guarantees consistently smooth ramp
- Visual (audible optional) indication of high voltage applied to DUT
- Discharge circuit provides for rapid discharge of DUT following shutdown
- Visual and audible indication of the following:
- Arc detection
- Excessive overcurrent flow
- Ground continuity test failure