|
|
ROD-L M100DC |
|
|
|
|
|
|
|
|
|
| MODEL M100DC |
 |
| DC Hipot Test Instrument |
 |
|
| Safe, Fast, Efficient |
 |
|
|
 |
Performs Dc dielectric withstand test up to 5.5 KVDC at 10 milliamperes (5mA standard) |
|
|
 |
Performs concurrent ground continuity test |
|
|
 |
ARC and OVERCURRENT detection circuits |
|
|
 |
Fast HV shutdown within 2 milliseconds |
|
|
 |
Fast discharge of Device Under Test (DUT) and other test standards |
|
|
| Production/Quality Assurance Testing |
|
|
 |
Fully automatic test cycle |
|
|
 |
Detects and indicates the following |
|
|
|
| » |
Test Ready (security ground chassis sensed) |
|
|
| » |
Test in Progress |
|
|
| » |
Test Passed |
|
|
| » |
Test Failed |
|
|
|
|
|
 |
Following test parameters easily preset via rear panel potentiometers |
|
|
|
| » |
Voltage ramp |
|
|
| » |
Test time |
|
|
| » |
Test voltage |
|
|
| » |
overcurrent trip point |
|
|
|
|
|
 |
Switch-selectable ramp and rest time combinations |
|
|
 |
Arc detection circuit |
|
|
 |
Overcurrent detection circuit |
|
|
 |
Solid state HV generation and control system for reliability |
|
|
 |
Switchable "burn" featured permits destructive testing by maintaining output voltage, following failure, to the DUT throughout duration of preset test time |
|
|
 |
Can be interconnected with M25 Ground Continuity Test Instrument or any ROD-L AC Hipot Test Instrument to simplify multiple testing requirements |
|
|
|
|
| Continuos Ground Continuity Monitoring |
|
|
 |
Monitors Ground Continuity of DUT before and during each Hipot Test |
|
|
 |
Senses for 0.5 ohm (standard), at approximately 1.5 VAC |
|
|
 |
Loss of safety ground connection during test causes high voltage to shut down with both audible and visual failure indication |
|
|
 |
Inhibits application of high voltage if the security chassis ground is not secure |
|
|
| Superior Safety Features |
|
|
 |
Front panel receptacle is rated to 8000 VAC |
|
|
 |
Front panel receptacle prevents direct operator exposure to high voltage during test |
|
|
 |
Shutdown within 2 milliseconds after dielectric breakdown, overcurrent or ground continuity fault |
|
|
 |
Electronically controlled output voltage ramp rate circuit guarantees consistently smooth ramp |
|
|
|
Visual (audible optional) indication of high voltage applied to DUT |
|
|
|
Discharge circuit provides for rapid discharge of DUT following shutdown |
|
|
|
Visual and audible indication of the following: |
|
|
|
| » |
Arc detection |
|
|
| » |
Excessive overcurrent flow |
|
|
| » |
Ground continuity test failure |
|
|
|
|
|
|
|
|
|