| MODEL M300RT |
 |
| Insulation Resistance Test Instrument |
 |
| Safe, Fast, Efficient |
 |
|
|
 |
Performs DC insulation resistance test - detects up to 4000 megaohms at 500 VDC |
|
|
 |
Test in compliance with UL, VDE, IDE, and MIL STD 202 test standards |
|
|
| Production/Quality Assurance Testing |
|
|
 |
Fully automatic test cycle |
|
|
 |
Detects and indicates the following: |
|
|
|
| » |
Test Ready (security chassis ground sensed) |
|
|
| » |
Test in Progress |
|
|
| » |
Test Passed |
|
|
| » |
Test Failed |
|
|
|
 |
Following test parameters easily preset via rear: |
|
|
|
| » |
Voltage ramp |
|
|
| » |
Test time |
|
|
| » |
Insulation resistance trip point |
|
|
|
 |
Switch-selectable ramp and test time combinations |
|
|
 |
Arc detection circuit |
|
|
 |
Solid state control circuitry for 2 millisecond shut down |
|
|
| Superior Safety Features |
|
|
 |
Shutdown within 2 milliseconds after insulation breakdown, or low insulation resistance |
|
|
 |
Electronically controlled output voltage ramp rate circuit guarantees consistently smooth ramp |
|
|
 |
Visual (audible optional) indication of high voltage applied to DUT |
|
|
 |
Visual and audible indication of insulation breakdown or low insulation resistance |
|
|
|